Diagnostics not functioning
The problem when the [Diagnostics] or [Machine Interface] is not working is when there is a problem with the AD Card. The issue can be caused by the drivers for the AD Card are not present, they have been corrupted or the AD Card itself has faulted. ...
Pedestal purge has low flow to pedestal
Check that there is no blockage or leaks to the purge tubing. If a blockage has occurred, check the pedestal support plate purge inlet. In the event that there is a leak, check that the silicone tubing connecting the pedestal purge O2 to the ...
Syringe pump cavitation
Syringe pump cavitation (air bubbles in syringe during operation) can be caused by the following reasons; Aspiration line is too long. (Tubing going to reservoir) Syringe pump speed set too fast for the syringe size. Slow down the pump speed to a ...
Z Depth requires adjusting
You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...
Replace Broken Pedestal
To install a new pedestal; Remove carousel Remove protective boot - rear of autosampler Remove 2 autosampler bay covers - left and right Unscrew old pedestal from pedestal support plate Screw new pedestal into pedestal support plate Repeat steps 3-1