Replace Broken Pedestal

Replace Broken Pedestal

To install a new pedestal;
  1. Remove carousel
  2. Remove protective boot - rear of autosampler
  3. Remove 2 autosampler bay covers - left and right
  4. Unscrew old pedestal from pedestal support plate
  5. Screw new pedestal into pedestal support plate
  6. Repeat steps 3-1

    • Related Articles

    • Pedestal purge has low flow to pedestal

      Check that there is no blockage or leaks to the purge tubing. If a blockage has occurred, check the pedestal support plate purge inlet. In the event that there is a leak, check that the silicone tubing connecting the pedestal purge O2 to the ...
    • Z axis and head unit targeting issue

      The Z motor pinion gear is loose and the Y timing belt require replacement/adjustment.