Calibration factor moved high
Very important that the operators change the Anhydrone at regular intervals. As the Anhydrone absorbs moisture, the occurrence that this is done is dependant on the dryness of the sample and also humidity that may enter the open tube combustion system. You can tell by looking at the Anhydrone if it is depleting or not is when it appears to begin “clumping” or the granules appear to be “fused” together.
The IR Cell lenses should be OK if the above has been adhered to. If the cell lenses are dirty, you would observe a drop in cell voltage. E.G. from 5vdc to around 3.5/4vdc.
Related Articles
Linearisation and Calibration
The linearisation is a set of values that should never change. The only case where this may change is when an IR emitter has faulted and the new heat signature from the new emitter is different from the previous emitter. As the emitter is not a ...
mircro switch for pistons and door malfunctioning
Has the lower sensor on the loader cylinder moved out of position? Could you check these 2 things please. Access to the sensor is from the rear panel directly behind the loader mechanism. Simply adjust the lower sensor on the cylinder until the LED ...
Turntable is out of alignment and keeps stopping at position 28.
The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder. Turn the monitor so ...