Sample was heated, but result not measurable and not useable.
Make sure that the continue analysis on unstable result is checked in the global analysis settings. The results that are marked with an “X” can be output by turning on the [ok] (Bottom right column at bottom of screen) from the [Print Import\Format Print] screen. This can be done from historics and sample entry screen.
Related Articles
Method setting not taking effect
Make sure to refresh the method by re-selecting it in the [Sample Entry] screen. When you hover over the method name with the mouse the method parameters will be displayed.
Analysis - Import Setup
Importing Your Samples There are only 9 available fields that can be imported into the program. The program only needs to set the formatting once, therefore once the formatting is complete the user will not be asked to format the file again. You can ...
Calibration factor moved high
Very important that the operators change the Anhydrone at regular intervals. As the Anhydrone absorbs moisture, the occurrence that this is done is dependant on the dryness of the sample and also humidity that may enter the open tube combustion ...
Error 5004 Samples Schemes
This error occurs when the system on startup is expecting to load a [Samples Scheme.json] file but this file does not exist or the system has the wrong file path information. To fix this error do the following: Check the [Configuration] screen to the ...
Z Depth requires adjusting
You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...