Z Depth requires adjusting

Z Depth requires adjusting

You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the Z limit in the [Targets] screen and adjust again if required. Note down the desired [Z Position] displayed on screen as this will be needed to change the targets json file. The targets scheme will then require adjustment. Open the C:\Labfit\AS3000\Schemes\Targets Schemes\"xxxx.json" file and not the "Targets Presistent Scheme" file. Locate all [Z Position] entries and change to the new Z depth noted above.
    • Related Articles

    • Z axis and head unit targeting issue

      The Z motor pinion gear is loose and the Y timing belt require replacement/adjustment.
    • Z Axis Constant Move Error

      Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
    • Turntable is out of alignment and keeps stopping at position 28.

      The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder. Turn the monitor so ...