Z Axis Constant Move Error
Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
Related Articles
Z axis and head unit targeting issue
The Z motor pinion gear is loose and the Y timing belt require replacement/adjustment.
Z Depth requires adjusting
You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...
How to change AS speed settings
Refer to page 89 of the user manual and set all speeds as shown in the tables for each axis in the [Hardware] screen. Click: Edit Settings Change values for each axis as shown on page 89. Click save Restart software The speed profile will now be set.
Diagnostics not functioning
The problem when the [Diagnostics] or [Machine Interface] is not working is when there is a problem with the AD Card. The issue can be caused by the drivers for the AD Card are not present, they have been corrupted or the AD Card itself has faulted. ...
Error 5004 Samples Schemes
This error occurs when the system on startup is expecting to load a [Samples Scheme.json] file but this file does not exist or the system has the wrong file path information. To fix this error do the following: Check the [Configuration] screen to the ...