Z Axis Constant Move Error

Z Axis Constant Move Error

Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
    • Related Articles

    • Z axis and head unit targeting issue

      The Z motor pinion gear is loose and the Y timing belt require replacement/adjustment.
    • Z Depth requires adjusting

      You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...
    • How to change AS speed settings

      Refer to page 89 of the user manual and set all speeds as shown in the tables for each axis in the [Hardware] screen. Click: Edit Settings Change values for each axis as shown on page 89. Click save Restart software The speed profile will now be set.
    • Error 5004 Samples Schemes

      This error occurs when the system on startup is expecting to load a [Samples Scheme.json] file but this file does not exist or the system has the wrong file path information. To fix this error do the following: Check the [Configuration] screen to the ...
    • CS2000 Error Writing to File

      Your IT department must fix the read/write permissions to your operating directory: Windows XP: C:\Program Files\CS2000 – Full access Windows 10 32 & 64 bit: C:\CS2000 – Full access Error occurs because the user account that is currently logged does ...