CS 12132 furnace not heating 2 pulses on proportional controller
If you see 2 pulses on the proportional controller this means that is has no incoming AC. Check incoming power has not tripped. Places to look are;- Switch Board 32A breaker
- Rear of CS1232 32A Breaker
- Check wiring inside orange power supply unit