CS 12132 furnace not heating 2 pulses on proportional controller

CS 12132 furnace not heating 2 pulses on proportional controller

If you see 2 pulses on the proportional controller this means that is has no incoming AC. Check incoming power has not tripped. Places to look are;
  1. Switch Board 32A breaker
  2. Rear of CS1232 32A Breaker
  3. Check wiring inside orange power supply unit


    • Related Articles

    • Controller not running

      The schemes do not have the correct information stored in the *Persistent Scheme.json file. Meticulous care needs to be taken to ensure that the system reads in the all the data correctly. In particular the following schemes are critical to the ...
    • Turntable is out of alignment and keeps stopping at position 28.

      The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder. Turn the monitor so ...