Controller not running

Controller not running

The schemes do not have the correct information stored in the *Persistent Scheme.json file. Meticulous care needs to be taken to ensure that the system reads in the all the data correctly. In particular the following schemes are critical to the system starting up correctly: Measurements, Samples, Data, Targets & Hardware. Each of the *Persistent Scheme must have a file that exists for it to load properly. E.G. if you have a Samples Scheme names "Test1.json" then the [Samples Persistent Scheme] must have only 1 entry called ["Test1.json"] with the quotes.
    • Related Articles

    • Error 5004 Samples Schemes

      This error occurs when the system on startup is expecting to load a [Samples Scheme.json] file but this file does not exist or the system has the wrong file path information. To fix this error do the following: Check the [Configuration] screen to the ...
    • CS 12132 furnace not heating 2 pulses on proportional controller

      If you see 2 pulses on the proportional controller this means that is has no incoming AC. Check incoming power has not tripped. Places to look are; Switch Board 32A breaker Rear of CS1232 32A Breaker Check wiring inside orange power supply unit
    • CS2000 Error Writing to File

      Your IT department must fix the read/write permissions to your operating directory: Windows XP: C:\Program Files\CS2000 – Full access Windows 10 32 & 64 bit: C:\CS2000 – Full access Error occurs because the user account that is currently logged does ...
    • Z Axis Constant Move Error

      Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
    • Z Depth requires adjusting

      You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...