Turntable is out of alignment and keeps stopping at position 28.

Turntable is out of alignment and keeps stopping at position 28.

The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder.
  1. Turn the monitor so that you can see the [Diagnostics] screen from where you will be manually moving the sensor.
  2. Remove air pressure from the CS analyser.
  3. Manually move the cylinder so that it is in the inner ring position.
  4. Loosen the sensor closest to the direction you moved the cylinder to achieve the inner ring position.
  5. Gently move the sensor on the cylinder until it goes off/on and then on/off (from the diagnostics screen). You need to find the point where sensing occurs. Mark the outer limit of this position using the sensor band as a guide.
  6. Repeat step 5 but for the other side of the sensor band. You should now have 2 marks on the cylinder.
  7. Move the sensor so that it is central to the 2 marks made from step 5 & 6 above.
  8. Tighten sensor band.

    • Related Articles

    • Replace Broken Pedestal

      To install a new pedestal; Remove carousel Remove protective boot - rear of autosampler Remove 2 autosampler bay covers - left and right Unscrew old pedestal from pedestal support plate Screw new pedestal into pedestal support plate Repeat steps 3-1
    • Z Depth requires adjusting

      You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...
    • Unable to output results

      Go to [Print Carousel] and then [Format Print]. Turn on the fields to be exported and set the column number for the delimiter position for that field. Next go to [Print to File] option. Click the [Browse] button and navigate to the directory where ...
    • mircro switch for pistons and door malfunctioning

      Has the lower sensor on the loader cylinder moved out of position? Could you check these 2 things please. Access to the sensor is from the rear panel directly behind the loader mechanism. Simply adjust the lower sensor on the cylinder until the LED ...
    • Analysis - Import Setup

      Importing Your Samples There are only 9 available fields that can be imported into the program. The program only needs to set the formatting once, therefore once the formatting is complete the user will not be asked to format the file again. You can ...