CS Series
Diagnostics not functioning
The problem when the [Diagnostics] or [Machine Interface] is not working is when there is a problem with the AD Card. The issue can be caused by the drivers for the AD Card are not present, they have been corrupted or the AD Card itself has faulted. ...
Pedestal purge has low flow to pedestal
Check that there is no blockage or leaks to the purge tubing. If a blockage has occurred, check the pedestal support plate purge inlet. In the event that there is a leak, check that the silicone tubing connecting the pedestal purge O2 to the ...
Replace Broken Pedestal
To install a new pedestal; Remove carousel Remove protective boot - rear of autosampler Remove 2 autosampler bay covers - left and right Unscrew old pedestal from pedestal support plate Screw new pedestal into pedestal support plate Repeat steps 3-1
mircro switch for pistons and door malfunctioning
Has the lower sensor on the loader cylinder moved out of position? Could you check these 2 things please. Access to the sensor is from the rear panel directly behind the loader mechanism. Simply adjust the lower sensor on the cylinder until the LED ...
CS 12132 furnace not heating 2 pulses on proportional controller
If you see 2 pulses on the proportional controller this means that is has no incoming AC. Check incoming power has not tripped. Places to look are; Switch Board 32A breaker Rear of CS1232 32A Breaker Check wiring inside orange power supply unit
Turntable is out of alignment and keeps stopping at position 28.
The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder. Turn the monitor so ...
Unable to output results
Go to [Print Carousel] and then [Format Print]. Turn on the fields to be exported and set the column number for the delimiter position for that field. Next go to [Print to File] option. Click the [Browse] button and navigate to the directory where ...
Method setting not taking effect
Make sure to refresh the method by re-selecting it in the [Sample Entry] screen. When you hover over the method name with the mouse the method parameters will be displayed.
Sample was heated, but result not measurable and not useable.
Make sure that the continue analysis on unstable result is checked in the global analysis settings. The results that are marked with an “X” can be output by turning on the [ok] (Bottom right column at bottom of screen) from the [Print Import\Format ...
Calibration factor moved high
Very important that the operators change the Anhydrone at regular intervals. As the Anhydrone absorbs moisture, the occurrence that this is done is dependant on the dryness of the sample and also humidity that may enter the open tube combustion ...
CS2000 Error Writing to File
Your IT department must fix the read/write permissions to your operating directory: Windows XP: C:\Program Files\CS2000 – Full access Windows 10 32 & 64 bit: C:\CS2000 – Full access Error occurs because the user account that is currently logged does ...