mircro switch for pistons and door malfunctioning

mircro switch for pistons and door malfunctioning

Has the lower sensor on the loader cylinder moved out of position? Could you check these 2 things please.  Access to the sensor is from the rear panel directly behind the loader mechanism. Simply adjust the lower sensor on the cylinder until the LED lights up. The best way to do this is to move the sensor so that the LED light goes off. Mark this location with a maker pen/texter/sharpie. Move the sensor band in the opposite direction until the LED lights and then goes off. Mark the sensor band location on the cylinder. Move the sensor band to the center of the 2 maks on the cylinder. Sensor should now be cetral between the 2 marks. This will ensure no "false positives are triggered by the sensor.


    • Related Articles

    • Z Axis Constant Move Error

      Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
    • Turntable is out of alignment and keeps stopping at position 28.

      The inner ring cylinder sensor requires adjustment. The inner ring cylinder sensor is on the edge of sensing. Adjust the sensor so that the sensing position is central to the pickup (located inside the cylinder) on the cylinder. Turn the monitor so ...
    • CS 12132 furnace not heating 2 pulses on proportional controller

      If you see 2 pulses on the proportional controller this means that is has no incoming AC. Check incoming power has not tripped. Places to look are; Switch Board 32A breaker Rear of CS1232 32A Breaker Check wiring inside orange power supply unit
    • Pedestal purge has low flow to pedestal

      Check that there is no blockage or leaks to the purge tubing. If a blockage has occurred, check the pedestal support plate purge inlet. In the event that there is a leak, check that the silicone tubing connecting the pedestal purge O2 to the ...