How to change AS speed settings
Refer to page 89 of the user manual and set all speeds as shown
in the tables for each axis in the [Hardware] screen. Click:
- Edit Settings
- Change values for each axis as shown on
page 89.
- Click save
- Restart software
The speed profile will now be set.
Related Articles
Z Axis Constant Move Error
Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
Z axis and head unit targeting issue
The Z motor pinion gear is loose and the Y timing belt require replacement/adjustment.
Syringe pump cavitation
Syringe pump cavitation (air bubbles in syringe during operation) can be caused by the following reasons; Aspiration line is too long. (Tubing going to reservoir) Syringe pump speed set too fast for the syringe size. Slow down the pump speed to a ...
Sample was heated, but result not measurable and not useable.
Make sure that the continue analysis on unstable result is checked in the global analysis settings. The results that are marked with an “X” can be output by turning on the [ok] (Bottom right column at bottom of screen) from the [Print Import\Format ...
Z Depth requires adjusting
You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...