Error 5004 Samples Schemes

Error 5004 Samples Schemes

This error occurs when the system on startup is expecting to load a [Samples Scheme.json] file but this file does not exist or the system has the wrong file path information. To fix this error do the following:
  1. Check the [Configuration] screen to the C:\Labfit\AS3000\System.ini. All values must match.
  2. Quit the AS3000 software.
  3. Locate and open the [Measurements Persistent Scheme.json] file in the C:\Labfit\AS3000\Schemes\Measurement Persistent Scheme directory.
  4. Compare the value between the inverted commas: e.g. "My Sample Scheme Name.json" to that of the file located in C:\Labfit\AS3000\Schemes\Samples Schemes. These file names must match each other in both files.
  5. Correct the value in the C:\Labfit\AS3000\Schemes\Measurement Persistent Scheme.json to that of the C:\Labfit\AS3000\Schemes\Samples Schemes.
  6. Save and close
  7. Restart the AS3000 software.

    • Related Articles

    • Controller not running

      The schemes do not have the correct information stored in the *Persistent Scheme.json file. Meticulous care needs to be taken to ensure that the system reads in the all the data correctly. In particular the following schemes are critical to the ...
    • Analysis - Import Setup

      Importing Your Samples There are only 9 available fields that can be imported into the program. The program only needs to set the formatting once, therefore once the formatting is complete the user will not be asked to format the file again. You can ...
    • CS2000 Error Writing to File

      Your IT department must fix the read/write permissions to your operating directory: Windows XP: C:\Program Files\CS2000 – Full access Windows 10 32 & 64 bit: C:\CS2000 – Full access Error occurs because the user account that is currently logged does ...
    • Z Axis Constant Move Error

      Adjust the Z Home sensor (Brass Ring) on probed holder. Make sure the magnet is aligned to within 5mm from the end or the front (where wires are) of the switch.
    • Z Depth requires adjusting

      You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...