Recent Articles
Pedestal purge has low flow to pedestal
Check that there is no blockage or leaks to the purge tubing. If a blockage has occurred, check the pedestal support plate purge inlet. In the event that there is a leak, check that the silicone tubing connecting the pedestal purge O2 to the ...
Syringe pump cavitation
Syringe pump cavitation (air bubbles in syringe during operation) can be caused by the following reasons; Aspiration line is too long. (Tubing going to reservoir) Syringe pump speed set too fast for the syringe size. Slow down the pump speed to a ...
Z Depth requires adjusting
You need to adjust the upper Z stop moving it upwards. Mark the bottom of the Z stop with a marker pen so that a visual representation can be observed when moving the Z stop. This will allow the probes to reach further into the sample cups. Test the ...
Replace Broken Pedestal
To install a new pedestal; Remove carousel Remove protective boot - rear of autosampler Remove 2 autosampler bay covers - left and right Unscrew old pedestal from pedestal support plate Screw new pedestal into pedestal support plate Repeat steps 3-1
mircro switch for pistons and door malfunctioning
Has the lower sensor on the loader cylinder moved out of position? Could you check these 2 things please. Access to the sensor is from the rear panel directly behind the loader mechanism. Simply adjust the lower sensor on the cylinder until the LED ...